Endress+Hauser TMT31-2087/0 Temperature Transmitter
- Supply Type
- Sample
- No
- Transport Mode
- Sea, Land, Air, Express, Others
- HS Code
Product Description
Product Details
| Brand | Endress+Hauser(E+H) |
| Model | TMT31-2087/0 |
| Type | Temperature Transmitter |
| Origin | Germany |
| Product Condition | New |
| Unit Price | $300 |
| Warranty | 12 months |
📌Overview
Endress+Hauser (E+H) TMT31-2087/0 Temperature Transmitter is a single-channel temperature transmitter. The temperature transmitter is mainly used to convert signals from Pt100/Pt1000 resistance temperature detectors (RTDs) or thermocouples (TCs) into standard analog outputs. TMT31-2087/0 supports two installation modes: installation inside a B-type terminal head or DIN rail mounting in a control cabinet.
✅Benefits
Installation and Wiring: TMT31-2087/0 uses push-in terminals for tool-free, quick wiring.
CvD Sensor-Transmitter Matching: TMT31-2087/0 uses Callendar-Van Dusen coefficients to optimize the overall measurement accuracy of the measuring point.
Sensor-Transmitter Matching (CvD): TMT31-2087/0 uses the Callendar-Van Dusen linearization function to adjust the accuracy of the entire measuring point and eliminate errors caused by individual sensor variations.
⚙️ Mechanical Installation Precautions
- Strict Torque Control: The maximum torque for the terminal head fixing screws is 1 Nm (Z2 screwdriver); the maximum torque for the terminal block screws is 0.35 Nm (Z1 screwdriver). Excessive torque may strip the threads or damage the housing/terminals, while insufficient torque may cause poor contact and loss of explosion-proof protection.
- Installation Orientation: When the TMT31-2087/0 transmitter is installed in the terminal head, power supply terminals 1 and 2 must face the cable entry. The sensor cable must pass through the central 7 mm hole. Do not pull the wires forcefully.
- Mandatory Explosion-Proof Requirement: After wiring is completed, the terminal head cover must be tightened and securely locked. Failure to tighten the cover will result in loss of Ex explosion-proof protection. Opening the cover while operating in hazardous areas is strictly prohibited.
- Do not apply excessive pressure to, drop, or impact the TMT31-2087/0 transmitter. When the terminal head is directly installed on the temperature measuring insert, avoid transferring mechanical stress to the transmitter body.
🔌Technical Specifications
- Measuring principle | Head transmitter
- Input | 1x Pt100 / Pt1000, TC
- Accuracy:
- (Pt100, Pt1000, -50...+250°C) ≤0,1 K
- (Pt100, Pt1000, -58...+482°F) ≤0,18 °F
- Weight:0.2g
🏢Contact Information