1.Technical data:
MODEL |
PH-2603 |
Measuring range |
pH: 0.00~14.00 mV: 0~ ±1999mV °C: 0~100°C EC: 0.00~19.99ms/cm CF: 0.0~199.9CF TDS: 10~19990ppm |
Resolution |
pH: 0.01pH mV: 1mV °C: 0.1°C EC: 0.01ms/cm CF: 0.1CF TDS: 10ppm |
Accuracy |
pH: ±0.01pH mV: ±1mV °C: ±0.4°C EC.CF.TDS: ±2%F.S |
Input impedance |
102 Ω |
Calibration |
1 or 2point |
Working temperature |
0~95°C |
Power |
DC9V/200mA or Rechargeable battery |
Operating conditions |
5~60°C,RH≤95%(Without magnetic interference around the meter & conspicuous centrum) |
Dimension |
208×200×76mm |
Weight |
510g |
2.Operation
PH MEASUREMENTS:
Turns on the PH electrode the temperature sensor and the power source. The DC/RB switch must choose DC Turn the meter ON by pressing the ON/OFF key.Remove the PH electrode protective cap, then submerge the PH electrode and the temperature probe into the sample to be tested.Stir gently and wait for the stability symbol. The display will show the PH value automatically compensated for temperature.After use, rinse the electrode with clean water, replace the electrode protective cap.
ORP MEASUREMENTS:
Oxidation Reduction Potential(ORP) measurements provide a quantification of the oxidizing or reducing power of the sample tested.Connect the ORP electrode to the PH connector.To enter the mV mode turn the instrument ON and press the mV key until the display changes to mV.Then submerge the ORP electrode into the sample to be tested and allow time for the reading to stabilize.
TEMPERATURE MEASUREMENTS:
Make sure the temperature probe is connected to the meter.Turn the instrument ON and press the key to select the temperature mode.Dip the temperature probe into the sample, allow the reading to stabilize and read the temperature value.
EC CF or TDS MEASUREMENTS:
Connect the EC electrode to the EC connector.To enter the EC CF or TDS mode turn the instrument ON and press the MODE key until the display changes to ms/cm CF or ppt.Then submerge the EC electrode into the sample to be tested and allow time for the reading to stabilize.
3.Calibration:
Pour a small quantity of PH6.86 and 4.01(or PH9.18)solution into clean beakers.For a particularly accurate calibration, it is advised to use two beakers for each buffer solution .the first is to be used for rinsing the electrode,the second is to be used for the calibration, in this way, the risks of contaminating the buffer solution are reduced to a minimum.Switch on the instrument. Press the PH key to display the PH measurement.Immerse the PH electrode and temperature probe into PH 6.86 buffer solution and then shake briefly.Regulate the trimmer PH7 found in the lower left position with a screwdriver until the buffer solution value corresponding to the measurement temperature is obtained.
1. Our service
Gemc are synonymous with quality, personalized service and a commitment to excellence. With our specialized rapid, diagnostic test technologies, high quality assurance, brand recognition, innovative research and development techniques, customer-centric design, an experienced management team, and an energetic sales force, Gemc continues to set the standards in product quality and reliability. We believe in continually improving our products and service, and making the business relationship a beneficial experience for both customers and employees.
Based on our core competence in rapid test technology, high quality assurance in products, international renowned brand name, innovative R&D, customer-centric product design as well as strong sales network, the experienced management team and energetic sales force of Gemc will lead our life to achieve peacefulness, harmony, and higher quality.
2. How to order our product
(1) Customers to contact to consult , and purchase our products
(2) Our trader make a perfessional form according to your product
(3) Inform the product you need after you confirmation of the product
Telephone: 86-371-60997506
Fax: 86-371-60997506
Website: http://www.gv-medic.com
Address: NO.20,Bosong Road, Zhengzhou, Henan ,China
Zip: 450001
City: Zhengzhou
Country: China (Mainland)